Table 56 Acquisition specifications — OMNI-M-PA3232
Frequency
Digitizing frequency 100 MHz (10 bits)
A-scan (acquisition
mode)
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Maximum pulsing rate Up to 10 kHz (C-scan)
Depth in material 59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
Display
Refresh rate A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Envelope (echo-dynamic
mode)
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
On internal clock 1 Hz to 10 kHz
External pace Yes
On encoder On 2 axes: from 1 step to 65536 steps
Table 57 Data specifications — OMNI-M-PA3232
Processing
Number of data points Up to 8000
Real-time averaging 2, 4, 8, and 16
Rectification RF, full wave, half wave+, half wave