DMTA-20015-01EN [U8778402], Rev. M, February 2017
Appendix N
190
Figure N-5 Fastening hole location and specifications
N.1 Acoustic Specifications
This section details the acoustic specifications of the pulser, receiver, and beam
formation, for both UT and PA modes (see Table 69 on page 190 and Table 70 on
page 191).
Table 69 Acoustic specifications — UT channel using UT connector
Pulser
Voltage 95 V, 175 V, and 340 V
Pulse width Adjustable from 30 ns to 1000 ns; resolution
of 2.5 ns
Fall time <10 ns
Pulse shape Negative square pulse
Output impedance <30 Ω
28 mm 162 mm
109 mm
47 mm 90 mm
M6, 7.9 mm deep
92 mm