10-19
10. Data Format Commands (P, G, R, W)
10.18 BLOCK B
BLOCK COMMAND PREFIX COMMAND SUFFIX (HEX)
B R,W 41
NOTE: R,W means “R” or “W”
This block consists of 17 items and 19 bytes (38 ASCII characters):
TABLE 10.18 R/W41 (BLOCK B)
*W412A202020010115030000002000000894040000
HEX BYTE NUMBER DEFAULT
COMMAND DESCRIPTION NUMBER OF VALUE
SUFFIX & ORDER BYTE
1E Recognition Charater 1 1 2A
1F Units of Measure 2, 3, 4 3 202020
20 Serial Delay 5 1 01
1A RS-485 Device Address 6 1 01
18 COM.PAR and MOdbUS 7 1 15
13 NUM.dLy (Alarm Delay) 8 1 33
12 AL.CNF and AL.MOdE 9 1 00
11 AL.CNF 10 1 00
10 SP.CNF 11 1 00
05 INPUt and Rd.SC.OF 12 1 00
0C CNt.by and dEC.Pt 13 1 00
16 OUt.CNF and Ad.RAtE 14 1 00
07 RDG.CNF 15 1 08
1C BUS.FMt 16 1 94
1B dAt.FMt 17 1 04
0E FILtER 18 1 00
0A INP.CNF 19 1 00

Table of Contents