USER'S MANUAL
General instructions for use
(Rev. 0) I-Max Touch (110-120V)
89
NOTE:
The I-Max Touch assumes that the digital sensor is ready: if this is not
the case, the following error message will be displayed:
D I G I T A L S E N S O R
I S N O T R E A D Y
Refer to the Manual of the Digital Acquisition System to correct the
situation.
To reset the message on the I-Max Touch, press key .
NOTE:
X-rays are emitted with a delay of two seconds from pressing the X-ray
button to allow the heating of the filament and the control of all set
parameters. Since the X-ray button is a "dead man’s switch", it must
be kept pressed until the end of the exposure.
3. Once the exposure is completed, the secondary collimator moves into
a back resting position, to allow the patient to come out.
The display will again show all the exposure values relating to the
exposure just completed.
NOTE:
If you want to carry out a new exposure, but the necessary waiting time
for the cooling of the anode hasn't yet passed, the display will show a
message indicating the time you still have to wait before carrying out the
new test:
T U B E C O O L I N G
P L E A S E W A I T x x x s
This time enables the X-ray tube's anode to cool down.
*
*
*