USER'S MANUAL
General instructions for use
I-Max Touch (110-120V) (Rev. 0)
94
You can pass from Normal Resolution (indicated by the letter "n") to
High Resolution (indicated by the letter "h"), by pressing key
"Examination Mode Selection" (10) and viceversa.
Pressing twice the key (10) the unit will return to
PANORAMIC STD position; the display shows:
C O N F I M P A N ?
> 0 < = S , T = N
Pess the "Patient Entrance" (6) key to confirm or the "Test" (5)
key to cancel the setting.
NOTE:
With the same image format, the scanning time is lower in Normal
Resolution; this allows you to give the patient a smaller dose, yet still
obtaining an image of sufficient quality for the orthodontic diagnostics,
albeit with a spatial resolution lower compared with that obtained from
High Resolution images.
5. Regulate the exposure parameters as required, using the pre-set
values or manual selection; the display will show the kV and mA
settings as per the following table.
Child
kV mA
Small 60 6
Medium 60 6
Large 60 6
Table 7
*