- PCME STACK 990 MANUAL: 57 -
Self Test Min. Pass
Level
Optimum
Level
Max. Pass
Level
Zero Check - <10 100
Span Check 9000 10000 11000
Short Circuit Check 8000 10000 15000
Contamination
Ring check
- <10 100*
* User programmable
Disabling Self Test
The self tests for each sensor can be enabled/disabled:
Select ‘Setup’ display
Menu route:
More Settings
Edit an existing device
Select device to edit
Self Tests Active
Yes / No
Select Save to save the settings and return to the main menu.
The alarm for each self test can be enabled/disabled:
Select ‘Setup’ display
Menu route:
More Settings
Edit an existing device
Select device to edit
Zero/Span alarm?
(Short circuit alarm?)
Contam Ring alarm?)
Yes / No
Select Save to save the settings and return to the main menu.
Adjusting the Contamination Ring Check Alarm Level
The influence of the contamination ring signal may vary for different installations.