SGM700 Digitizer
Main Menu -continue-
5.2 CAL code (Calibration counter).
The SGM700 has a CAL code inside. CAL code is the number of times the
calibration is changed. When an indicator gets certified this number will be written
on the device and is used by the controlling agency to see if the settings aren’t
changed after sealing.
To check out the CAL code Press key 3 <2 sec. to go into the Main menu.
Press key 1 <2sec. 2x to the dCAL menu
and press key 2 <2sec. to enter.
The following screen will show the actual CAL code.
Press Key 3 <2 sec. to go back to the main menu.
Press key 3 <2 sec. to go back to the main weigher display.
31