Configuration Menu -CAL- -continue-
Normally, the deadload is zero, but it is possible to change the line position if
there is weight on the scale. To do so, edit the actual weigh value to the new
known value.
When the new dead load is set the follo-
wing screen is visible:
Configuration Menu -tCL-
6.16 - - - tCL Theoretic calibration
In this menu, the Theoretic calibration settings can be set. Here you can set a
calibration without using a refrence weight. For this you only need the specificati-
on sheets of the used loadcells. Options are:
84
tCL 1 Maximum load loadcells
tCL 2 Sensivity loadcell 1
tCL 3 Sensivity loadcell 2
tCL 4 Sensivity loadcell 3
tCL 5 Sensivity loadcell 4