Advanced settings
34
SPECTRO
Selection Choice - setting
limit
Initial settings
Tracer Gas Helium 4 /
Helium 3 /
Hydrogen
Helium 4
Filament selected 1 / 2 1
Filament Off / On On
Filament Status 0 - 100 % 100 %
Calibrated Leak Tracer Gas Helium 4 /
Helium 3 /
Hydrogen
Helium 4
Type Internal / External Internal
Unit mbar·l/s /
Torr·l/s /
Pa·m
3
/s
mbar·l/s
Leak Value - Refer to certificate
delivered with the de-
tector
Calibration valve Closed / open Closed
Loss Per Year (%) 0 - 99 6
Reference Temperature of (°C) 0 - 99 23
Temperature Coefficient of (%/°c) 0.0 - 9.9 3.0
Year - Refer to certificate
delivered with the de-
tector
Internal Temperature (°C) (If type = internal)
External Temperature (°C) (If type = external)
--
MAINTENANCE
Selection Choice - Setting
limit
Initial settings
Detector -20
Timers Detector - 20
Filament 1 Timer - 20
Reset Timer Function launching - -
Filament 2 Timer - 20
Reset Timer Function launching - -
Calibrated leak - To be set
Cycle Counter Cycle Counter - 0
Time interval 1·10
+19
- 1 5·10
5
Reset counter Function launching - -
Primary Pump Timer (h) - 20
Time interval (h) Wet Model 0 - 99999 8600
Dry Model 0 - 99999 17200
Reset Timer Function launching - -
Secondary Pump 1 Timer (h) 20
Time interval (h) 8600
Reset Timer Function launching - -
Speed (rpm) - -
Detector Information Access to Detector general information - -