Telemetry Functions at the Information Center
Telemetry Functions at the Information Center & TeleMon
9-13
Suppress
SpO
2
Inops
with NBP
Enable/disable the SpO
2
algorithm to detect NBP running
and suppress sending technical
alarms from the transceiver for
60 seconds.
WarningWarning
If you measure SpO
2
on a limb
that has an inflated NBP cuff,
a non-pulsatile SpO
2
technical
alarm can occur. If the
monitor is configured to
suppress this alarm, there can
be a delay of up to 60 seconds
in indicating critical patient
status, such as sudden pulse
loss or hypoxia.
enable
disable
enable
Pleth Wave Enable/disable the transmission
of the Pleth wave and its
subsequent display to the
Information Center. For
Continuous mode only.
enable
disable
Note—When enabled, during
6-lead monitoring, the Pleth
wave will replace the Vb
wave in the Patient Window.
disable
(Pleth wave
is not
displayed.)
Pulse Enable/disable the transmission
of the Pulse rate and its
subsequent display to the
Information Center. For
Continuous mode only.
enable
disable
disable
(Pulse rate is
not
displayed.)
Unit Settings - Telemetry Setup
Control Function Settings
Factory
Default

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue TRx and is the answer not in the manual?

Philips IntelliVue TRx Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue TRx
CategoryTransceiver
LanguageEnglish

Related product manuals