Draft - 1 Aug 08
Telemetry Functions at the Information Center
9-14 Telemetry Functions at the Information Center & TeleMon
SpO
2
Alarm Turn SpO
2
alarms on/off at the
Information Center.
enable (on)
disable (off)
enable
SpO
2
Limits
High
Increment/decrement SpO
2
high
alarm limit by 1 (in %).
Limit maximum is 100. Limit
minimum is 51 (adult) or 31
(pediatric). High and low
limit must be at least 1%
apart.
100 (adult,
pediatric)
SpO
2
Limits
Low
Increment/decrement SpO
2
low
alarm limit by 1 (in %).
Limit maximum is 99. Limit
minimum is 50 (adult) or 30
(pediatric). High and low
limit must be at least 1%
apart.
90 (adult,
pediatric)
Unit Settings - Telemetry Setup
Control Function Settings
Factory
Default
Unit Settings - Default Leads
Control Function Settings
Factory
Default
Leadset
Allowed
Set the unit default leadset type. All, EASI, No EASI, 3 wire All
3-wire Set the unit default lead. I, II, III II
5-wire, ECG1 Set the unit default lead. I, II, III, MCL, AVR, AVL,
AVF, V
II
5-wire, ECG2 Set the unit default lead. I, II, III, MCL, AVR, AVL,
AVF, V
V