The behaviour is factory-programmed. If another switching function
is wished, the pump can be reprogrammed in the
‘Relay’
menu.
The variable to be signalled for the current output can be selected
in the
‘ANALOGUE OUTPUT’
menu.
The current output plus relay can be retrofitted and operates once
it is plugged into the board.
Electrical interface
for current output
Data Value Unit
Open circuit voltage: 8 V
Current range: 4 ... 20 mA
Ripple, max.: 80 μA ss
Load, max.: 250 Ω
for semiconductor switch ("relay"):
Data Value Unit
Max. residual voltage at I
off max
= 1 µA 0.4 V
Pacing pulse duration, approx. 100 ms
To pin VDE cable Contact Relay
1 yellow "+" Current
output
4 Green “-" Current
output
3 white N/C (normally closed)
or
N/O (normally open)
Relay
2 brown C (common) Relay
Fig. 28: Assignment on the cable
Identity code c
Fig. 29: Assignment on the pump