MP3-PL Therapy Beam Analyzer T41050
Operating Manual – Installation of the System for Measurement
Vertical mounting refer to Figure 15
Mount the holder T21004.1.004 on the C-slider
with the two M4x16 screws (4, Figure 15).
Mount the linear detector holder with the Pin-
Point chambers on the holder T21004.1.004 us-
ing the two M4x40 screws (2, Figure 15).
Figure 15: Vertical mounting of the linear detector holder with 24 PinPoint chambers
1 Linear detector holder T21003.1.001
2 Screw M4x40 (2x)
3 Holder T21004.1.004
4 Screw M4x16 (2x)