5 Error Messages and Troubleshooting
If problems occur, please check the installation as
outlined in section 2. The following list of possible
problems may be a useful guideline for trouble-
shooting.
For detailed information on the individual compo-
nents, please refer to the corresponding user man-
uals.
For control pendant operating instructions and trou-
bleshooting tips, please refer to the user manual of
the control unit and the control pendant.
When contacting our Service team, please give
the full serial number of your device.
1. The Power LED at the
TBA CONTROL UNIT is
not illuminated.
Check whether the power cord is connected.
Is the control unit turned on?
Check the fuses in the cold appliance socket on the back of the device.
2. No information on the TBA
CONTROL PENDANT
display.
Is the control pendant connected?
When the control pendant is connected to the junction box:
Is the control cable T43164/K1 properly connected between control unit
and junction box?
Is the POWER LED at the control unit illuminated? Refer to 1.
3. An axis does not respond
when you push one of the
direction buttons.
If the control pendant reports a message, please refer to the user manual
of the control pendant.
Check the axis limits. If upper and lower limit are identical, the axis cannot
be moved. For configuration of the limits, refer to the user manual of the
control pendant.