MP3-PL Therapy Beam Analyzer T41050
Service Manual Cleaning
D850.131.05/01 en
Service Manual
WARNING
Improper execution of service tasks.
Hazards to Persons! Equipment Damage!
Ensure that the service tasks described below
are only performed by specialized staff.
In addition, ensure that all other service tasks are
only performed by PTW-Freiburg or by a person
authorized by PTW-Freiburg.
8 Cleaning
General Information on Cleaning
WARNING
Liquids have entered the product.
Shock Hazard, Erroneous Measurement!
The device must be thoroughly inspected by
service personnel before being used again.
CAUTION
Disinfection of the surface with disinfectants on a
phenol base or peroxide compounds.
Equipment Damage!
Do not use disinfectants on a phenol base or
peroxide compounds to disinfect the surface.
The product cannot be sterilized or completely dis-
infected.
Clean the product simply by wiping it down with a
dry or slightly moistened cloth. Use water or a mild
soap solution.
Alternatively, you can use a common disinfection
agent for medical instruments which is specified in
the list of disinfectants of your national hygiene
institute (e.g., VAH in Germany). If your country
does not provide such a list, we recommend either
a solution on an aldehyde base or a solution with a
quaternary ammonium compound.