MP3-PL Therapy Beam Analyzer T41050
Service Manual Cleaning
52
D850.131.05/01 en
Device Surfaces
First of all, disconnect the MP3-PL system from
the power line, before attempting to clean the
device surfaces.
Wipe the device down. Do not let liquid enter the
devices. Do not apply spray to the devices.
Connection Cables
Clean the cables as follows:
Disconnect the cables from the devices (always
pull on the connectors, not on the cables!).
Wipe the cables down.
Do not immerse the cables in liquids!
Connectors
Do not immerse the connectors in liquids! Do not let
liquid enter the connectors.
The connectors must not become dirty. Clean the
connectors only with a soft brush. Do not blow into
the connector.
Perspex Tank
The Perspex tank is drained by pumping the water
into the SCANLIFT water reservoir.
The inclination required to completely drain the
Perspex tank can be attained by adjusting one of
the leg levelers of the positioning device according-
ly. Do not tilt the tank manually.
The supporting profile and the angle are adjusted
by PTW-Freiburg. Do not remove them.
For cleaning, you can also use a mild cleaning
agent suitable for plastic material.
After cleaning, thoroughly remove all remains of the
cleaning agent.
You should coat the Perspex tank for storage with
the delivered cover.
Moving Mechanism
The shafts and toothed wheels of the moving
mechanism should be lubricated with the supplied
grease at intervals of 1 to 2 years or when neces-
sary (use small amount of grease only!).