MP3-PL Therapy Beam Analyzer T41050
Service Manual – Preventive Maintenance
9 Preventive Maintenance
Below, you will find a list of the maintenance
measures and inspections necessary for the prod-
uct.
Check before each use
Before each use, visually inspect the device and
the cables for signs of mechanical damage.
If damages or malfunctions are identified, the
device must be repaired before it is used again.
Safety inspections
Only regularly inspected devices are fail-safe. To
preserve the functional security and operational
safety, a safety inspection is to be executed at
regular intervals according to national regulations.
These inspections must only be performed by inde-
pendent persons with adequate training and experi-
ence.
It is recommended to execute the safety inspections
every 2 years.
Technical Inspection of the Measuring System
The system must be inspected at regular intervals
according to national regulations. It is recommend-
ed to perform a technical inspection of the measur-
ing system every 2 years at PTW-Freiburg or a
qualified calibration laboratory.
Inspections must also be performed after repairs or
after each influence which may have changed the
behavior of the device.