MP3-PL Therapy Beam Analyzer T41050
Operating Manual – Installation of the System for Measurement
3.3.2 Bragg Peak Chamber Type 34073
Attach the plug holder T4316/U161 to the Per-
spex tank.
If required, connect the ionization chamber to
the extension cable T2954/K2-6.
Clamp the plug connection of detector cable and
extension cable in the plug holder T4316/U161.
Connect the detector cable or the extension
cable to detector input labeled CHANNEL 1 on
the rear side of the MULTIDOS.
3.4 Mounting and Connecting
the Reference Detector
Figure 21: Mounting the reference detector
1 Holder for reference detector
2 Reference detector
3 Clamping screw (2x)
As reference detector you can use the following
chambers with the appropriate holders:
Bragg Peak chamber Type 34073 with holder
T41029.1.210,
Bragg Peak chamber Type 34070 with holder
T41029.1.260 and
Thin-Window Bragg Peak chamber Type 34080
with holder T41029.1.260