MP3-PL Therapy Beam Analyzer T41050
Operating Manual Installation of the System for Measurement
D850.131.05/01 en
Pumping from the Perspex Tank into SCANLIFT
Water Reservoir
To start the pumping process from the Perspex
tank into SCANLIFT water reservoir briefly press
once the WATER key on the SCANLIFT con-
trol pendant.
To stop the pumping process press the STOP
key.
3.10.3 Draining the System
CAUTION
Improper handling.
Equipment damage!
After use, the Perspex tank should be drained to
prevent deformation of the walls and undue
stress on the seams.
The Perspex tank may only be filled with water
continuously for 12 hours maximum in order to
avoid a damage of the Perspex tank due to the
hygroscopic character of PMMA.
The Perspex tank should not be filled again for
further 12 hours. This allows the water that was
incorporated in the PMMA to emit.
CAUTION
Improper handling.
Equipment damage!
The inclination required to completely drain the
Perspex tank can be attained by adjusting one of
the leg levelers of the positioning device accord-
ingly. Do not tilt the tank manually.
First empty the Perspex tank by draining the
water from the Perspex tank into the SCANLIFT
water reservoir.
To start the emptying process briefly press
once the WATER key on the SCANLIFT
control pendant.
To stop the emptying process press the
STOP key.
Move the SCANLIFT to a suitable drain and use
the screw plug underneath the SCANLIFT to
drain the water.