Page 89
Chapter 6 Parameter Listing and Descriptions
Application Parameters
Point 1 / Point 2
P1-P100, P2-P1100 Mode of Operation
This parameter sets the mode of operation, when in run mode, and can be set to
one of the following:
Display shows the distance from the
transducer face to the surface/interface.
Display shows how full the vessel is.
Display shows how empty the vessel is.
Display shows the average level of point 1
and point 2.
Display shows the differential between
point 1 and point 2.
* If transducer 2 is enabled
P1-P101, P2-P1101 Xducer (Transducer)
This parameter should be set to the transducer being used with the unit.
Transducers prefixed with “dB” are air transducers and those prefixed with “VT”
are submersible sludge transducers. This parameter can be set to one of the
following:
For point 2 only. Disables point 2 (default).
Transducer is a dB3. Range 0.125 to 3.00 metres
Transducer is a dB6. Range 0.3 to 6.00 metres
Transducer is a dB10. Range 0.3 to 10.00 metres
Transducer is a dB15. Range 0.5 to 15.00 metres
Transducer is a dB25. Range 0.6 to 25.00 metres
Transducer is a dB40. Range 1.2 to 40.00 metres
Transducer is a VT10. Range 0.3 to 10.00 metres