Secon 4
Integra 3™ Product Overview
15
Research Instruments Ltd
4
Manipulator
Fine Control
X, Y and Z movement from one lever
(X = side to side, Y = front to back, Z = up and down)
Sub-micron resoluon
0.75mm X and Y travel, 5mm Z travel
Manipulator
Coarse Control
X and Y movement
10 micron resoluon
4mm X and Y travel
PL3 Toolholders
Pipee locaon
Pipee angle adjustment 16 to 43 degrees
Heang Systems
(conguraon dependent)
Heated Metal Insert: 16mm
25mm
ITO Heated Glass Insert
Thermosafe Air Heang System
Temperature Controller
Accuracy: beer than ±0.1°C when calibrated against
known reference.
Displayed resoluon: 0.1°C
Setpoint Temperature Range: 30-45°C
Displays 3.5” LCD touch screen display for temperature control
Connecvity
USB Type B socket for connecon to PC running RI Viewer
or RI Witness soware (where used). Refer to soware
manual for further informaon.
Connected PC to be compliant with IEC 60950-1
XY Stage
Aluminium Heated Stage Plate
40mm travel in X and Y (25 and 33mm per turn)
Syringes Choice of SAS (air), SAS-SE (air) and SOS (oil)
Dimensions
Footprint (not including microscope)
Width 56cm, depth 38cm
Weight
Maximum weight of microscope mounted components
(excluding front and rear adaptors): 10.4kg
Actual weight is conguraon dependent
Mains Power Adaptor
Input: 100-240VAC, 50-60Hz, Max. 1.8A, Class I
Output: 12VDC, Min. 11A (132W)
Operang Range
Temperature: 15°C (59°F) to 40°C (104°F)
Humidity: 15% to 85% RH (non condensing)
Micropipee
Compability
Compable with Research Instruments micropipees
(or other industry standard 1mm OD micropipees of
equivalent specicaon).
Power
Supply
Adaptor
Required
Power
Rang
Serial
Number