EasyManua.ls Logo

Research Instruments INTEGRA 3 - Integra 3 Specification Table

Default Icon
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Secon 4
Integra 3™ Product Overview
15
Research Instruments Ltd
4
 
Manipulator
Fine Control
X, Y and Z movement from one lever
(X = side to side, Y = front to back, Z = up and down)
Sub-micron resoluon
0.75mm X and Y travel, 5mm Z travel
Manipulator
Coarse Control
X and Y movement
10 micron resoluon
4mm X and Y travel
PL3 Toolholders
Pipee locaon
Pipee angle adjustment 16 to 43 degrees
Heang Systems
(conguraon dependent)
Heated Metal Insert: 16mm
25mm
ITO Heated Glass Insert
Thermosafe Air Heang System
Temperature Controller
Accuracy: beer than ±0.1°C when calibrated against
known reference.
Displayed resoluon: 0.1°C
Setpoint Temperature Range: 30-45°C
Displays 3.5” LCD touch screen display for temperature control
Connecvity
USB Type B socket for connecon to PC running RI Viewer
or RI Witness soware (where used). Refer to soware
manual for further informaon.
Connected PC to be compliant with IEC 60950-1
XY Stage
Aluminium Heated Stage Plate
40mm travel in X and Y (25 and 33mm per turn)
Syringes Choice of SAS (air), SAS-SE (air) and SOS (oil)
Dimensions
Footprint (not including microscope)
Width 56cm, depth 38cm
Weight
Maximum weight of microscope mounted components
(excluding front and rear adaptors): 10.4kg
Actual weight is conguraon dependent
Mains Power Adaptor
Input: 100-240VAC, 50-60Hz, Max. 1.8A, Class I
Output: 12VDC, Min. 11A (132W)
Operang Range
Temperature: 15°C (59°F) to 40°C (104°F)
Humidity: 15% to 85% RH (non condensing)
Micropipee
Compability
Compable with Research Instruments micropipees
(or other industry standard 1mm OD micropipees of
equivalent specicaon).

Power
Supply
Adaptor
Required
Power
Rang
Serial
Number

Table of Contents