Secon 7
Integra 3™ Temperature Control
Research Instruments Ltd
7
When Integra 3 is powered up, the display briey shows RI and the
product logo. Aer start up the screen shows the actual temperatures
of the devices connected. The heang systems are connually
monitored for safe operaon.
Various heang opons are available for the Integra 3.
A brief descripon and how to idenfy each opon is detailed below.
Refer to the Integra 3 Installaon and Service Manual (6-54-701IM) for
further informaon about changing the heang opon (mode).
To ensure embryo safety, the temperature setpoint should
be set to maintain the specimen at the correct temperature.
See ‘Adjusng Specimen Temperature’.
In the event of a temperature sensor malfuncon, the setpoint
temperature may be exceeded and the surface may become hot
to the touch. It is recommended that temperatures are checked
on a regular basis.
The Heated Stage Plate and, when ed, the Heated Metal Insert
are shown on the main screen. The setpoint of each can be changed
individually, however, in pracce the setpoint of each will normally be
set to a similar value.
The Heated Stage Plate and when ed, the Heated Glass Insert
are shown on the main screen. The setpoint of each can be changed
individually. In pracce the setpoint temperature of the Heated Glass
Insert will normally be set slightly higher than the Heated Metal Stage.
The Heated Stage Plate and Heated Metal Insert can be used with the
Thermosafe Air Heang System. The setpoint of each can be changed
individually, however in pracce the setpoint of each will normally be
set to a similar value.