Secon 10
Integra 3™ Troubleshoong
52
Research Instruments Ltd
10



 






E01
HEATED
STAGE HEATING
FAILURE
Low
Heated Stage Plate
or Thermosafe Air
Heang System is
not able to heat.
Alarm acvated
2 minutes aer
power on if there
is less than 1.0°C
temperature rise
between 1 and
2 minutes aer
power on. If the
temperature at 1
minute is already
within ±2.5°C from
the setpoint the
test is omied.
Heated
Stage Plate
controller
power is set
to 0% unl
the
Integra 3
is re-started.
Switch o the
Integra 3 and
verify that the
Heated Stage
Plate cables
and connector
are secure and
undamaged.
Possible heater
or controller
fault.
E02
THERMOSAFE
HEATING
FAILURE
Low
Thermosafe
Air Heang
System
controller
power is set
to 0% and fan
speed is set
to 0% unl
the Integra 3
is re-started.
Switch o the
Integra 3 and
verify that the
Thermosafe
Air Heang
System cables
and connectors
are secure and
undamaged.
Possible heater
or controller
fault.
The following tables on pages 52 to 62 lists Alarm Condion Code
numbers, displayed Error Message, Priority, Fault Descripon, Alarm
Acons and a Soluon to the fault.
