9
To minimise the risk of injury:
• Never point the pipee towards yourself or anyone else
• Always make sure the pipee is securely mounted in the holder
before use
• When removing a pipee from the holder, be sure to hold the
pipee between your ngers as the holder p is loosened
• Release the pressure before loosening the MPH p
A self adhesive warning label is supplied, which should be placed
in a prominent posion as close as possible to the device when in
use.
High pressure can be generated in the system when using small
diameter micropipees, viscous uids, or if a micropipee
becomes blocked. This may cause micropipees to shoot out of
the holder.
US Federal Law restricts this device for sale to or on
order of a physician. This device is intended for use for Intra
Cytoplasmic Spermatozoa Injecon (ICSI). All other uses must be
for invesgaonal purposes or in the research laboratory only.
Do not re-use micropipees - reuse poses a high risk of cross
contaminaon.
Handle glass pipees with care.
Release the pressure in the syringe before releasing the pipee
from its holder.