Secon 10
Integra 3™ Troubleshoong
51
Research Instruments Ltd
10
  
Heated Stage Insert or
Heated Stage Plate too
hot or too cold
Insucient seling
me
Aer switching on the system should be
le for the display to show a constant
temperature before use
Dierence in
temperature between
specimen and plate
surface
Adjust the setpoint to allow for the
dierence (see page 34)
Temperature setpoint
incorrect
Touch the screen on the icon corresponding
to the aected heang system. Change if
necessary then press 

  
Objecves hit Heated
Stage Insert when
changing objecves or
adjusng focus
Objecves are focused
too high
Refocus on the top surface of a Petri dish
The specimen is too
close to the edge of
the hole in the insert,
causing the objecve
to foul the side of the
hole
Posion the specimen as close as possible to
the centre of the hole in the insert
Short working distance
objecve ed
Please contact RI for more informaon