Integra 3™ Safety Warnings
5
Research Instruments Ltd
3
Secon 3


  

RF emissions
CISPR 11
Group 1
The Integra 3 uses RF energy only
for its internal funcon. Therefore,
its RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Integra 3 is suitable for use
in all establishments other than
domesc, and may be used in
domesc establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domesc
purposes, provided the following
warning is heeded.
This equipment/system
is intended for use by healthcare
professionals only. This equipment/
system may cause radio interference
or may disrupt the operaon
of nearby equipment. It may
be necessary to take migaon
measures, such as re-orienng or
relocang the Integra 3 Or shielding
the locaon.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuaons/
icker emissions
IEC 61000-3-3
Complies
The Integra 3 is intended for use in the electromagnec environment
specied below.
The customer or the user of the Integra 3 should assure that it is used in
such an environment.