EasyManua.ls Logo

Research Instruments INTEGRA 3 - Syringes

Default Icon
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Secon 7
Integra 3™ Temperature Control

Research Instruments Ltd
7
Repeat procedure for Adjusng Specimen (Setpoint) Temperature
aer performing calibraon.
x
x
x
x
Figure 7-2 Figure 7-3
 

Place the probe of a calibrated thermometer in good thermal contact
with the surface. Simply touching the probe on the surface is not
adequate. Use a purpose-made surface probe or use thermal transfer
paste. Products sold for computer heatsinks are suitable, and RI can also
supply suitable materials.
When the system is ed with the Heated Stage Plate and Heated
Metal Insert, these must be calibrated independently.
Aer switching on,  to allow the temperature to
stabilise before calibrang.
Recommended posions for temperature measurement during
calibraon of heated plates are shown below for both metal
(Figure 7-2) and glass (Figure 7-3) heated stage opons.
Temperature calibraon can be performed on each of the heang opons
listed at the start of the Temperature Control secon. The calibraon
screen is accessed by pressing the  icon on the main screen, then
the  icon.
The displayed temperature is adjusted using the  or 
icons unl the value matches that of the thermometer used to perform
the calibraon. Following calibraon and aer returning to the main
screen, it is advisable to leave the probe in posion for a short while in
order to verify that temperatures are as expected. The recommended
method of calibraon is described below.
Place
probes
here

Table of Contents