Secon 9
Syringes

Research Instruments Ltd
9
Figure 9-9
Figure 9-10
1. Stand the SOS on its end as shown below. Fill the syringe supplied
with oil, unscrew the cap from the ll port and screw the syringe
onto the Oil Filling Port.
2. Press the syringe plunger slowly unl the level of oil in the cylinder is
just below the p. Look closely for any air bubbles in the oil. Any that
are seen can be dislodged by a short, sharp movement of the lling
syringe and should be allowed to rise to the top of the oil.
3. When you are certain there are no air bubbles, connue pressing
the syringe plunger to ll the cylinder and the tubing unl a small
amount of oil emerges from the p of the Micropipee Holder.

RI recommends sterile ltered paran oil.
Syringes and caps must be screwed ontothe ll port with gentle
nger pressure only. Overghtening may damage the ngs.