Secon 10
Integra 3™ Troubleshoong

Research Instruments Ltd
10
  
The specimen
falls o the
holding pipee/
unable to get
pressure to
inject
Tip of the
micropipee
holder has not
been ghtened
properly
Tighten the p if necessary
Micropipee p
could be blocked
Replace the micropipee
There may be
an air leak in
the injecon/
aspiraon system
Check the connecon of the syringe tubing
at each end. If loose, cut 10mm from the
end of the tubing and reconnect. Check the
tubing for damage along length and replace
if necessary. Replace the O-ring in the
micropipee holder p (see Figure 11-1 on
page 66) if worn out. Replace the O-ring in
the SAS air syringe (see Figure 11-3 on
page 68)
Poor control
from SAS Syringe
Check for:
- Air leaks as above
- Blocked pipee
Poor control
from SOS Syringe
Check for:
- Air bubbles
- Blocked pipee
- Using correct medium and oil eg a sterile
ltered paran oil
- Oil leaks (treat as for air leaks)
Unable to
aspirate/inject
Has the specimen
dried up?
Heated Stage warmth can dry up an
insuciently protected specimen. A generous
layer of mineral oil over the specimen will
protect it from drying out.
Are the media you
are using within
their shelf life?
Some media and PVP deteriorate if stored
past their shelf life or if they have not been
kept refrigerated during storage or transit.