Roche Diagnostics
110 Instructions for Use · Version 10.0
7 Measurement 9180 Electrolyte Analyzer
Measuring procedure
The analyzer will display Thank You! and a brief countdown will begin. After the
analysis is finished, the test results will be displayed and printed out.
Figure 7-3
q
Values that are higher or lower than the programmed normal range are indicated by an
arrow pointing up or down.
q
If the Na
+
result is above 180 mmol/L (above 169.5 mmol/L for blood/serum) or below
95 mmol/L (below 89.6 mmol/L for blood/serum), the Li
+
value cannot be derived.
If an additional sample report is desired or the automatic sample report is turned off,
the results may be printed out by following these steps:
o Press NO.
o The prompt PRINT FUNCTIONS? appears.
o Press YES.
o The prompt Print last Sample Report? is displayed.
o Press YES.
o The measurement results will be displayed, and the report will print out.
o The analyzer returns to READY.
Direct ISE
o Press NO until QC/STD/DIALYSATE/URINE SAMPLE? appears; press YES
and then NO until Standard Sample? is displayed. Press YES to confirm.
q
Standard Mode always reports direct ISE values and is not affected by correlation factors
or the setting of the QCC service code.
Urine samples
Before measuring urine, dilute the sample with Urine Diluent in a ratio of 1 part
urine to 2 parts Urine Diluent (e.g., 1 mL urine and 2 mL Urine Diluent).
Thoroughly mix the sample and analyze in the urine mode.
Na K Cl
143.9 4.93 103.1
ROCHE 9180
ELEKTROLYT ISE
NA-K-CL
HOSPITAL WEST
Name: ..........
..........
Sample: STANDARD
Sample No.:28
Na=143.9 mmol/L
K = 4.93 mmol/L
Cl=103.1 mmol/L