Roche Diagnostics
58 Instructions for Use · Version 10.0
4 Theoretical foundation 9180 Electrolyte Analyzer
Electrode specifications
Potassium electrode
The potassium electrode is a membrane electrode used for in vitro diagnostic
measurement of potassium ions present in fluid samples. It is designated with a K
+
marking on the top surface of the housing.
Construction
Figure 4-4 Potassium electrode
Use and care
Potassium electrodes are manufactured for use in 9180 Electrolyte Analyzers.
Proper care should be used in handling and storage of the electrode. Never use strong
or abrasive cleaners such as alcohol or amphyl on the electrode since these will
damage the plastic housing.
Store the electrode in a clean, dry place only after the electrode has been cleaned and
rinsed with distilled water and dried with a lint-free cloth.
The o-ring seal should be installed in the electrode during storage.
Electrode housing:
transparent acrylic plastic
Measuring membrane:
potassium ion selective
Electrolyte chamber:
airtight, filled with electrolyte solution for K
+
electrodes
Pin connector:
silver, silver chloride (Ag / AgCl)
A Inner electrode E Ion-selective electrode
B Sample path F Electrolyte chamber
C Marking G Connecting pin
D O-ring seal H Marking - serial number - 5 digit lot number