Roche Diagnostics
96 Instructions for Use · Version 10.0
6 Installation and shutdown 9180 Electrolyte Analyzer
Shutdown
o Repeat the same procedure for the rear pump winding.
Figure 6-25
o Press YES. The prompt Do You Wish to DELETE All Data? will appear.
o By pressing YES, all QC values and statistics will be deleted and the sample
number is reset to 0. QC and normal ranges, correlation factors, printer settings,
date and time and SnapPak volume are set to default values. Also, all service codes
are deactivated.
It is recommended that all data is deleted if the analyzer is used again at a later
date.
o Press NO to retain all data.
u Clearing all data (p. 166).
6. Turning the analyzer "off"
o The prompt Shutdown Complete Turn Power Off will appear.
o Turn off the analyzer.
o Unplug the power cord from the receptacle.
o Close the front door.
o Clean all external surfaces of the analyzer as well as the areas accessible through
the main door and sample door.
u Clean sample probe and fill port (p. 135).
q
Clean electrical contacts with a clean, dry or moist cloth. Do not use Deproteinizer or
other disinfectants on electrical contacts.
o Shutdown is complete.
o Use the original packaging when transporting the analyzer.