Roche Diagnostics
Instructions for Use · Version 10.0 45
9180 Electrolyte Analyzer 3 Specifications
Environmental parameters
Solutions
SnapPak
CAUTION
Incorrect SnapPak storage conditions may lead to incorrect results
The performance of the SnapPak is not guaranteed when the SnapPak is stored at
temperatures outside the recommended range.
r Do not freeze the SnapPak. Always store the SnapPak according to the specified
storage conditions.
Storage and transportation conditions in original packaging:
Consider the following reagents:
Standard A For calibration of sodium, potassium, chloride, calcium, and lithium in the
9180 Electrolyte Analyzer.
Standard B For calibration of sodium, potassium, chloride, calcium, and lithium in the
9180 Electrolyte Analyzer.
Standard C For calibration of sodium, potassium, chloride, calcium, and lithium in the
9180 Electrolyte Analyzer.
Reference Solution A salt bridge for calibration and measurement in the 9180 Electrolyte Analyzer.
Storage temperature 15 to 30 °C < 24 months
Transportation temperature 0 to 45 °C
Stability:
Expiration date and lot number are printed on each
container label
On-board stability:
14 weeks
Table 3-15
Stability:
Expiration date and lot number are printed on each
container label
On-board stability:
14 weeks
Table 3-16
Stability:
Expiration date and lot number are printed on each
container label
On-board stability:
14 weeks
Table 3-17
Stability:
Expiration date and lot number are printed on each
container label
On-board stability:
14 weeks
Table 3-18