EasyManua.ls Logo

Roche 9180 - Potassium Electrode Not Calibrated; Testing the Sample Sensor; Testing the Snappak Sensor

Roche 9180
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics
160 Instructions for Use · Version 10.0
10 Troubleshooting 9180 Electrolyte Analyzer
Service functions
Potassium electrode not calibrated
If the potassium electrode cannot be calibrated after an electrode exchange, then a
serum or plasma sample should be introduced as an external sample.
o Call up the procedure for testing electrodes.
u Testing the electrodes (p. 159).
o Press NO until Test External Sample? is displayed. Press YES.
Prepare a serum or plasma sample and introduce the sample:
o The prompt Open Sample Door Introduce Sample appears. Introduce the
sample.
o The sample is aspirated until the prompt Wipe Probe/Close Sample Door is
displayed.
o Test the sample for 15 minutes.
o Repeat the procedure a second time.
Testing the sample sensor
The dry sample sensor should read between 80-120 units. Aspirating clear fluid
should increase this value by at least 40 units.
o Press NO until Test Sample Sensor? is displayed.
o Press YES and the analyzer will display the current sample sensor reading:
Dry: 80-120: 99
YES=Pump/NO=Exit
o Press YES to introduce an external sample through the probe. An example of the
analyzer display after this step would be:
External: 236
Pump: NO=OFF
o Press NO to exit and after a short Please Wait message, the prompt
Test SnapPak Sensor? will appear.
Testing the SnapPak sensor
o Press NO until Test SnapPak Sensor? is displayed.
o Press YES and the current SnapPak sensor status will be displayed:
SnapPak: I
I=In O=Out
o "I" on the display indicates that the SnapPak is inserted properly, "O" means that
the SnapPak either is not installed or not inserted correctly.
q
Allow 10 seconds for the analyzer to detect the SnapPak.
q
If the SnapPak has been removed during this test, a calibration should be performed to
prime all fluid lines.

Table of Contents

Related product manuals