9180 Electrolyte Analyzer 3 Specifications
Performance parameters
Material: Pooled Human Serum, n=80
Material: Urine, n=80
Parameter Mean S
wr
(CV %) S
dd
(CV %) S
T
(CV %)
Sodium 138.8 0.30 0.22 0.36 0.28 0.47 0.34
Potassium 4.49 0.034 0.75 0.041 0.92 0.051 1.13
Chloride 106.8 0.18 0.17 1.00 0.93 1.24 1.16
Calcium 1.19 0.007 0.55 0.031 2.64 0.039 3.29
Lithium 0.17 0.011 6.19 0.015 8.40 0.023 13.28
Table 3-8
Parameter Mean S
wr
(CV %) S
dd
(CV %) S
T
(CV %)
Sodium 51.5 1.98 3.84 3.06 5.94 3.65 7.08
Potassium 48.4 0.65 1.34 0.97 2.00 1.11 2.29
Chloride 85.9 0.53 0.62 0.66 0.76 0.99 1.16
Calcium N/A
Lithium N/A