102 Instructions for Use · Version 10.0
7 Measurement 9180 Electrolyte Analyzer
Preanalytics
Plasma samples older than 1 hour have to be re-centrifuged in order to remove fibrin
clumps that may have formed.
Each laboratory should determine which sample containers should be used with the
analyzer.
Aqueous solutions
Aqueous samples, such as Standard A, have to be measured in the Standard Mode.
q
Standard Mode always reports direct ISE values, and is not affected by correlation factors
or the setting of the QCC service code.
u Service codes (p. 156).