Deleting data
Deleting data
Clearing sample count
To reset the sample count to 0, perform the following steps:
o Open the READY display and press NO until OPERATOR FUNCTIONS?
appears. Confirm by pressing YES.
o Press NO until Reset Sample Number? appears. Press YES.
o When the prompt Are you sure? appears, press YES again.
Clearing all data
To reset the analyzer software to the default values, press the YES and NO keys
simultaneously and shut down and restart the analyzer. In addition, this function can
be performed after decommissioning the analyzer
All QC values and statistics will be deleted and the sample number will be set to 0. QC
and normal values, correlation factors, printer settings and date and time will be set
to default values. All service codes will be deactivated. It is recommended that all data
is deleted if the analyzer is used again at a later date.