Roche Diagnostics
Instructions for Use · Version 10.0 95
9180 Electrolyte Analyzer 6 Installation and shutdown
Shutdown
5. Installing the relief clamps and removing the pump windings
The display will prompt All 5 Relief Clamps Inserted?
o Install the red relief clamps making sure that they are snapped securely into
position.
Figure 6-24
The relief clamps can be installed at the following positions:
O A
O B
O C
O V
O R
o Press YES.
q
Do not insert the solenoid relief clamps with the SnapPak in place.
o The prompt Pump Windings Relieved? appears.
o Grasp the front winding close to the pump roller and gently pull it off the roller.