9180 Electrolyte Analyzer 3 Specifications
Sample throughput
Sample throughput
Measurement time:
Below 75 seconds per sample.
Additional printout takes up to 20 seconds
Sample volumes
95 μL typical
Sample types
Whole blood, serum, plasma, urine, QC material
Calibrations
1-point calibration with each measurement
2-point calibration every 4 hours