9180 Electrolyte Analyzer 9 Maintenance
Unscheduled maintenance
NOTICE
Use of non-approved consumables may null and void the warranty
Use of consumables that were not produced by Roche, including but not limited to
electrodes, calibration and cleaning solutions, QC-material and related accessories, may
result in malfunction of the analyzer and may render the warranty null and void.
r Only use approved consumables when working with the analyzer. For the full list of
approved consumables, see Available accessories and consumables (p. 176).
Replacing electrodes
o Slide the measuring chamber forward until it locks into the forward position.
o Unclamp the left electrode holder by moving the clamp forward.
Figure 9-9
o Remove the used electrode from the measuring chamber.
Figure 9-10