9180 Electrolyte Analyzer 6 Installation and shutdown
Shutdown
Figure 6-20
2. Installing the transport housing
o The analyzer will prompt Transport Ref Housing Installed?
o Pull the electrode holder forward and remove the reference electrode assembly.
o Carefully unscrew the reference electrode from the reference electrode housing.
o Temporarily place the reference electrode on a clean, soft cloth.
o Pour the remaining reference solution in the reference housing into a transport
housing.
o Carefully screw the reference electrode into the filled transport housing.
o The second transport housing is screwed into the reference housing.
o Place the reference housing back into the electrode holder and move the lever
backwards, making sure that all electrodes are seated properly. The electrode
holder remains in the forward position.
Figure 6-21
o Press YES.