Troubleshooting
128
Error Description/Solution
Error 4: Meter too warm
or too cold
Solution
Turn the meter off and unplug the power supply unit. Make sure that
the ambient temperature falls within acceptable limits (+18-32°C)
and wait for a few minutes until the meter has reached ambient tem-
perature.
Error 5: Voltage of
handheld battery pack
too low
No more tests can be performed.
Solution
Insert a new handheld battery pack as described on page 19 or
Connect the meter to the power supply unit to recharge the
handheld battery pack.
Error 6: Batteries in
meter
The meter may not be operated with batteries.
Solution
Insert the correct handheld battery pack for the meter.
Error 7: Result memory
full
Appears first, when only 10 more results can be stored. If result
memory is completely full, no more tests can be run.
Solution
Check the meter settings for data handling as described on page 44.
If you have enabled usage of a computer, you now can or must
transfer data to the specified host.
If you are not using a computer/host system, disable this setting.
Then, with each new test performed, the oldest data will be
deleted
Error
4
Error
No meas. left
5
Error
Battery
6
Error
Memory Full!
7

Table of Contents

Other manuals for Roche cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas h 232 and is the answer not in the manual?

Roche cobas h 232 Specifications

General IconGeneral
BrandRoche
Modelcobas h 232
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals