R&S NRP Remote Control - Commands
1144.1400.12 6.35 E-3
DISPlay[:WINDow][1..4]:RESolution[?] 1 | 0.1 | 0.01 | 0.001
Specifies the number of significant places of the mantissa when the measurement result is displayed on
a linear scale. The R&S NRP tries to adjust the averaging filter range so that the display noise
decreases with increasing resolution. The values of the transfer parameter have the following meaning:
Parameter Linear scale of measurement result Logarithmic scale of measurement result
1 1 significant place No decimal place
0.1 2 significant places 1 decimal place
0.01 3 significant places 2 decimal places
0.001 4 significant places 3 decimal places
Note: Automatic filter ranging must be switched on
(SENSe[1..4]:AVERage:COUNt:AUTO ON). DISP[1..8]:RES has an
effect on the setting SENS[1..4]:AVER:COUN:AUTO:RES of the sensors
used in the measurement function (CALC[1..8]:MATH). However, the
resolution setting of the sensors does not affect the resolution of values in
the display.
This command corresponds to the
<resolution> option of the high-level measurement commands
CONFigure, FETCH, READ and MEASure.
*RST value: 0.01
Error messages:
-127
"Invalid numeric data": An invalid number was specified for the resolution.
Example:
DISP:WIND1:RES 0.1
With a measured value of 123.456789 W, the first two places are free of noise and thus 120 W is
displayed.
DISPlay[:WINDow][1..4][:STATe][?] ON | OFF
Opens or closes a window. When a window is closed the other windows are automatically expanded so
that the free space is utilized. Likewise, open windows become smaller when a new one is opened.
*RST value: ON (Window 1)
OFF (Window 2,3,4)