Trying Out the Instrument
R&S
®
RTO
37Getting Started 1316.0833.02 ─ 10
8. Press the SLOPE key to toggle the trigger slope and watch the waveform and the
"Trigger" settings.
4.2 Acquiring Data
Although you are using only one input channel, you can acquire the calibration data
using three different arithmetic methods and display the results in separate waveforms.
1. Press the ACQUISITION key on the front panel (in the HORIZONTAL area).
2. In the "Acquisition" tab of the "Horizontal" dialog box, enable all three waveforms
("Wfm1", "Wfm2", "Wfm3").
3. For Wfm1, select the "Decimation" type Sample and the "Wfm Arithmetic" Off.
4. For Wfm2, select the "Wfm Arithmetic" Envelope. The "Decimation" type is auto-
matically set to Peak detect to display the correct envelope waveform.
5. For Wfm3, select the "Decimation" type High Res and the "Wfm Arithmetic" Aver-
age.
6. Close the "Horizontal" dialog box.
Acquiring Data