EasyManua.ls Logo

Rohde & Schwarz R&S RTO Series - Displaying the Waveform History

Rohde & Schwarz R&S RTO Series
133 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Trying Out the Instrument
R&S
®
RTO
49Getting Started 1316.0833.02 ─ 10
Tip: You can turn the Navigation knob to change the zoom factor while holding the
waveform.
4. Release the finger when the waveform segement of interest is visible in the zoom.
5. Tap "Keep" to convert the fingertip zoom into a standard zoom diagram.
4.6 Displaying the Waveform History
During a continuous acquisition, the acquired data is stored in the memory and the cur-
rent acquisition is shown on the display. When the acquisition was stopped and a new
acquisition is started with RUN CONT or RUN ×SINGLE, the memory is cleared and
written anew. The history accesses and displays the samples that were saved before
the current acquisition.
In the following example, you will acquire 10 waveforms, then display the 3 most recent
waveforms.
1. Press the ACQUISITION key on the front panel (in the HORIZONTAL area) to
open the "Horizontal" settings dialog box.
2. Set the "Average count" to 10 to perform 10 waveform acquisitions.
3. Close the "Horizontal" dialog box.
Displaying the Waveform History

Table of Contents