Trying Out the Instrument
R&S
®
RTO
63Getting Started 1316.0833.02 ─ 10
To determine the search conditions you can use the results of the measurement exam-
ple described in chapter 4.7.3, "Performing and Configuring the Quick Measurement",
on page 54.
To configure a search for overshoots
1. Press the SEARCH key on the front panel (ANALYSIS area).
2. Select C1 as the "Source".
3. Define the search criteria: Enable "Window".
4. In the "Window" tab, define the search conditions:
a) As the "Upper level", enter the result of the "High" level measurement.
b) As the "Lower level", enter the result of the "Low" level measurement.
c) As the "Vertical condition", select Exit to find values that are outside the range
defined by the high and low levels.
5. Select "Enable" to start the continuous search on the acquired data.
6. Close the "Search" dialog box.
7. In the "Search Results" box, select "Show search zoom windows".
The acquisition stops, and the detected overshoots of the last acquisition are listed
in the search result table. The last result that was found is displayed in the search
zoom window. Vertical lines indicate the time values for which a result was found.