Reference Manual
00809-0100-4664, Rev BA
January 2010
E-11
Rosemount 8712
ENDRESS AND HAUSER
SENSORS
Connect coil drive and electrode cables as shown in Figure E-9.
Endress and Hauser
Promag 10/50/53/55
H/P/WS Sensors
Figure E-9. Wiring Diagram for
Endress and Hauser Sensors
and Rosemount 8712
Table E-7. Endress and Hauser Sensor Wiring Connections
Rosemount 8712 Endress and Hauser Sensors
141
2 42
17
18 5
19 7
4
57 37
42 41
ROSEMOUNT 8712
TRANSMITTER
ENDRESS AND HAUSER
SENSORS
This is a pulsed DC magnetic flowmeter.
Do not connect AC power to the sensor
or to terminals 1 and 2 of the
transmitter, or replacement of the
electronics board will be necessary.