Reference Manual
00809-0100-4664, Rev BA
January 2010
Rosemount 8712
E-28
KENT SENSORS Connect coil drive and electrode cables as shown in Figure E-26.
Veriflux VTC Sensor to
8712 Transmitter
Figure E-26. Wiring Diagram for
Kent Veriflux VTC Sensor and
Rosemount 8712
Table E-24. Kent Veriflux VTC Sensor Wiring Connections
Rosemount 8712 Kent Veriflux VTC Sensors
15
2 2
SCR OUT
17 SCR OUT
18 SIG1
19 SIG2
Electrode Connections
Coil Connections
ROSEMOUNT 8712
TRANSMITTER
KENT VERIFLUX VTC
SENSOR
1SCR OUT
2
3 SIG 1
4 SIG 2
5
6
1
2–
5+
6 SCR OUT
Fuse
This is a pulsed DC magnetic flowmeter.
Do not connect AC power to the sensor
or to terminals 1 and 2 of the
transmitter, or replacement of the
electronics board will be necessary.