EasyManua.ls Logo

Rosemount 8712 - Page 168

Rosemount 8712
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Manual
00809-0100-4664, Rev BA
January 2010
Rosemount 8712
E-14
Endress and Hauser
Generic Sensor to
Rosemount 8712
Transmitter
Connect coil drive and electrode cables as shown in Figure E-12.
Figure E-12. Wiring Diagram for
Endress and Hauser Sensors
and Rosemount 8712
Table E-10. Endress and Hauser Generic Sensor Wiring Connections
Rosemount 8712
Endress and Hauser
Generic Sensors
141
2 42
14
17 4
18 5
19 7
41
41
42
42
5
7
ROSEMOUNT 8712
TRANSMITTER
ENDRESS AND HAUSER
GENERIC SENSORS
This is a pulsed DC magnetic flowmeter.
Do not connect AC power to the sensor
or to terminals 1 and 2 of the
transmitter, or replacement of the
electronics board will be necessary.