Rosemount Model 8732C Integral Mount Magnetic Flowmeter System
2-10
An inclined installation, as shown in Figure 2-8, is an acceptable
installation technique. This tends to keep the cross-sectional area full.
NOTE
As shown in Figure 2-8, avoid declining installations where back
pressure is inadequate to ensure that the flowtube remains full.
The electrodes in the Model 8711 are properly oriented when the top
of the flowtube is either vertical or horizontal, as shown in Figure .
Avoid any mounting orientation that positions the top of the flowtube
at 45° from the vertical or horizontal position.
Model 8711 Mounting Position.
Flow
B
8705-0005A01E, 0005A01F
FIGURE 2-8. Incline or Decline Orientation.
Flow
45° Electrode Plane
45° Electrode Plane
8711-8711E01A, 8711F01A